Explainability

A trained classifier gives you a class and a probability, and nothing about which part of the recording moved it there. Attribution answers that second question: it assigns every sample of every lead a contribution to one output class. That much is generic. What makes it useful for ECG is what comes next — collapsing the map into a distribution over the twelve leads, and averaging it over R-peak-aligned beats so the result can be read against the cardiac cycle. A heatmap over a 5000-sample twelve-lead tensor is not a claim anyone can check. "This model draws 10% of its attribution from the P-wave region, against the 25% that region's width would earn by chance" is.

from deepecgkit.explainability import ECGExplainer

explainer = ECGExplainer.from_checkpoint("runs/my-run/checkpoints/best-epoch.pt", sampling_rate=500.0)
result = explainer.explain(signals)

result.values            # (samples, leads, time) attribution, the input's own shape
result.signal            # the signal the model saw, same shape
result.lead_importance   # (samples, leads), each row sums to one
result.target            # (samples,) the class each row explains
result.beat_profile      # (leads, window) averaged over detected beats
result.pqrst_fractions   # {"P": 0.10, "QRS": 0.52, "T": 0.36, "baseline": 0.02}
result.pqrst_enrichment  # {"P": 0.41, "QRS": 4.18, "T": 0.96, "baseline": 0.08}
result.crop_offset       # samples dropped from the front to fit the model
deepecg explain --checkpoint runs/<run-directory> -d ptbxl \
    --method integrated-gradients -o explanations/

From the command line the data is prepared the way the checkpoint was trained, and the beat windows are placed at the sampling rate the model ends up seeing, so neither needs stating; see deepecg explain for what --sampling-rate means with a recording file.

Where the attribution comes from

The gradient is taken with respect to the input, holding the weights fixed. Attribution therefore comes back in the input's own layout, per lead and per timestep, which is what every summary below depends on.

inputs = signals.clone().detach().requires_grad_(True)
scores = model(inputs).gather(1, target.view(-1, 1)).squeeze(1)
gradients = torch.autograd.grad(scores.sum(), inputs)[0]

Two consequences worth knowing. Attribution runs the model in evaluation mode but not under torch.no_grad(), unlike feature extraction — it enables gradients itself, so calling it from inside an evaluation loop that has already entered no_grad works. And it uses torch.autograd.grad rather than backward(), so no parameter gradient is ever populated and a live optimiser cannot be corrupted by explaining a model mid-training.

Choosing a method

Method What it answers Cost per recording
saliency How much would this logit move if this sample moved slightly? one forward, one backward
integrated-gradients How much of the logit gap against a silent baseline does this sample own? ig_steps forwards and backwards
occlusion How much does the logit fall if this window is blanked out? one forward per window per lead

Saliency is the default because it is the only one cheap enough to run over a whole split. Integrated gradients is the one to quote, because it satisfies the completeness axiom: its attributions sum to the difference between the model's score on the signal and on the baseline, so each value is a share of a real quantity rather than a relative heat. Occlusion asks a different question entirely and does not depend on the gradient at all, which makes it a useful second opinion when a gradient result looks suspicious.

Saliency returns magnitude by default; integrated gradients and occlusion return signed values, where positive means the sample supported the prediction.

Which lead drives the prediction

Each lead is scored by its mean absolute attribution and the scores are normalised to sum to one, so lead_importance is a share rather than a magnitude. Magnitude rather than sign, because evidence against a class is still evidence the lead was read; a signed sum would let a lead with large opposing contributions cancel to near zero and read as ignored.

from deepecgkit.explainability import plot_lead_importance

plot_lead_importance(result.lead_importance, leads=result.lead_names, save_path="leads.png")

Twelve-lead data is labelled with the standard PTB-XL ordering automatically. A single-lead model returns [1.0], which is correct rather than degenerate — with one lead there is nowhere else for the evidence to be — but it tells you nothing, so do not report it.

Where in the beat the model looks

R-peaks are detected with a Pan-Tompkins style energy detector, one window is cut per beat, and the windows are averaged. Detection runs on the signal, never on the attribution: the cardiac cycle is a property of the recording, so anchoring to the attribution would make the window positions depend on the method under test.

The default beat window runs from 250 ms before the R-peak to 500 ms after it, and the regions are those a cardiologist would name:

Region Window relative to R Share of the beat window
P −250 to −50 ms 25%
QRS −50 to +50 ms 12.5%
T +100 to +400 ms 37.5%
baseline the remainder 25%

pqrst_fractions reports each region's share of the total attribution mass. Those shares are taken against the whole beat, not against the three named regions, so they do not sum to one — the remainder is the isoelectric baseline, and renormalising it away would make a model that attends mostly to the flat line look like one that attends to the P wave.

pqrst_enrichment divides each share by the region's share of the window's duration. One means the region received exactly the attention its width would earn by chance, above one means the model concentrates there, below one means it avoids it. This is the number the P-wave claim turns on: a P share of 0.10 sounds small until it is read against 0.25, at which point it says the model attends to the P-wave region at roughly 40% of chance.

!!! note "Attribution explains the model, not the patient"
    A high QRS enrichment means this network's decision depended on QRS samples. It is not
    evidence that the QRS complex is diagnostic, and it is not a clinical finding. Two models
    trained on the same data will give different attributions and both will be faithful to
    themselves.

!!! warning "Set the sampling rate"
    `sampling_rate` is what places the P, QRS and T windows around each detected R-peak. In
    Python nothing can check it against an array you pass, so giving 500 for a 100 Hz
    recording puts the QRS window where the T wave is and every fraction it reports will be
    wrong. The `deepecg explain` command takes the rate from the checkpoint and the dataset
    instead.

Visualisation

from deepecgkit.explainability import plot_attribution, plot_beat_profile

plot_attribution(result.signal[0], result.values[0], sampling_rate=500.0,
                 leads=result.lead_names, save_path="signal.png")
plot_beat_profile(result.beat_profile, result.beat_times_ms, save_path="beats.png")

These take (leads, time), matching the model input and the attribution tensor. The older plot_ecg_signals in Evaluation was written for (time, leads) and accepts either layout; the new plots take (leads, time) only, the convention used everywhere else in the package. Omitting save_path calls plt.show(), which will block in a script.

Model constraints

The same bounds that govern feature extraction govern attribution, because attribution runs the same forward pass. One thing differs: a recording above an architecture's ceiling is now centre-cropped for you and a RuntimeWarning says so, rather than failing inside the network. A recording below the minimum still raises, because cropping cannot make a signal longer.

The authoritative bounds live in the model registry, declared next to each architecture's @register_model decorator, and can be queried as deepecgkit.registry.get_model_info(name)["min_input_length"] / ["max_input_length"] (None meaning no floor beyond 1, or no ceiling). The table below is a convenience copy.

Model Min samples Max samples Centre-cropped above max
afmodel 7971 at recording_length=30, 1393 at 6 or 10 unbounded n/a
convnext-v2 29 unbounded n/a
crnn 16 unbounded n/a
deep-res-cnn 127 unbounded n/a
dualnet 8 5000 yes
fcn-wang 2 unbounded n/a
gru 1 unbounded n/a
inception-time 1 unbounded n/a
kanres 1 unbounded n/a
kanres-deep 11 unbounded n/a
lstm 1 unbounded n/a
mamba 50 25049 yes
medformer 50 5009 yes
resnet 1 unbounded n/a
resnet-wang 1 unbounded n/a
se-resnet 1 unbounded n/a
simple-cnn 8 unbounded n/a
tcn 1 unbounded n/a
transformer 1 5000 yes
xresnet 17 unbounded n/a
xresnet1d-benchmark 1 unbounded n/a

After a crop, result.values is shorter than the signal you passed and crop_offset gives the number of samples removed from the front, so sample i of the attribution is sample i + crop_offset of your recording. The discarded head and tail were not attributed zero; they were not attributed at all. Pass time_offset=result.crop_offset / sampling_rate to plot_attribution to keep the time axis anchored to the original recording.

Every model bakes its lead count into its first layer, so a lead-count mismatch raises before the forward pass rather than producing a plausible but meaningless result.

Cost

Measured on resnet at twelve leads and 5000 samples, on CPU: one forward and backward pass is about 74 ms. Saliency costs one of those. Integrated gradients at 64 steps costs about 4.7 s per recording. Occlusion at window=100, stride=50 is 1188 forward passes, about 13.5 s. Use --max-samples deliberately; it defaults to 8 for this reason.

mamba is far worse than the table suggests, because its selective scan runs a Python loop over the sequence — roughly 4000 sequential autograd nodes at its maximum length. Integrated gradients on mamba at full length is impractical.

Known issues

  • Gradient saliency measures sensitivity at a single point. It is noisy, it saturates wherever the network does, and it reacts to sharp local structure such as the QRS complex whether or not the decision depended on it.
  • Occlusion attribution bleeds one window width past any sharp boundary in the model's receptive field, because a window straddling the boundary perturbs samples on both sides and the single resulting score change is spread across all of it. Narrow the window to sharpen the result.
  • The zero baseline is a modelling assumption, not a neutral fact. It is the right choice for z-scored input, where zero is the per-lead mean, and a poor one for raw millivolts, where blanking to zero injects a step the model reacts to.
  • R-peak detection is a filtered-derivative threshold detector, not a clinically validated one, and it has not been checked against an annotated database. On noisy or low-amplitude leads it misses beats, and the profile is then an average over fewer beats than you assume. beat_count reports how many survived.
  • Atrial fibrillation is the case where detection is hardest and the claim matters most: RR intervals are irregular and a false peak on a T wave misaligns every window. Check beat_count against the recording length before quoting a P-wave fraction.
  • Fixed windows around each R-peak do not follow heart rate. Ectopic beats, wide QRS complexes and marked rate variation will all put real waveform outside the window it belongs to.
  • An untrained xresnet attributes only through its identity path, because xresnet1d.py zero-inits the residual BatchNorm weight. Attributions from a randomly initialised model of that architecture are degenerate rather than merely uninformative.
  • Values are not comparable across methods or across architectures. Never put a saliency map and an occlusion map on the same colour scale.
  • Medformer builds a cross_channel_attn block that forward never invokes, so 5.5% of its parameters receive no gradient and contribute to no attribution. This is the same defect noted in Feature Extraction.